Impact Ventilator

Circuits

• Adult: >25kg:

• Pedi ≤ 25kg

• NO VENT <5kg

Minute ventilation:

• 6-8 L/min

Ideal Body Weight:

• 5 Ft = 50 Kg

• Add 2.2 Kg for each inch above 5 feet

Mode: (AC, SIMV, PS, CPAP)

Inpsire - Vol/P

*Defaults to AC-V

BMP (Range 1-60 BPM)

*Defaults to 12

**Default alarms = 25/8

I:E vs Tiu (inspiratory timer - Range 0.0-3 sec)

Defaults to 1:2.5 & 1 sec Ti

BIPAP

• mode = CPAP, NPPV

• PIP/PEEP - press & hold for pressure support mode

• select PS # (i.e. 10PS, PEEP 5 = CPAP 10/5)

CPAP

• mode = CPAP, NPPV

• set PEEP

PIP (range 0-80 cm H20)

*Goal < 35

**Default alarms = 35/10 V, 20/10 P)

PEEP (range 0-25cm H20) *+5 = Physiologic PEEP

(Keep same Delta = Inspiration P)